Module 6 — Graded Quiz (Summative)
20 questions
Question 1 — Question 1
R1-A has been operating normally for several months. This morning, the pump-down from ~950 mbar to 1 mbar takes the usual ~90 seconds. Below 1 mbar, the pump-down slows dramatically and stalls at 0.30 mbar after 25 minutes. The pump was tested independently at its inlet and reaches 0.008 mbar — within specification. A rate-of-rise test shows a decreasing rate (0.08 to 0.02 mbar/min over 10 minutes). The operator also observes visible oil mist escaping from the R1-FLT-EXH exhaust outlet. What is the most likely cause of the performance degradation?
Question 2 — Question 2
A technician proposes replacing R1-P-RP (oil-sealed rotary vane pump) with a diaphragm pump to eliminate all oil contamination risk on R1-A. Using R1-A's normal operating parameters (base pressure 0.05 mbar, chamber volume pumped from ~950 mbar), interpret whether this proposal is viable.
Question 3 — Question 3
On R1-A, the operator closes R1-V-ISO and performs a rate-of-rise test. The rate is constant at 0.04 mbar/min. The pump was independently verified at 0.009 mbar (within spec). All chamber O-rings were replaced last week and appear to be in good condition. A colleague suggests: "Just get a bigger pump — more pumping speed will fix it." Given that base pressure scales inversely with effective pumping speed for a given gas load, interpret why this suggestion does not address the underlying problem.
Question 4 — Question 4
R1-A is in VENTED state (~950 mbar). A student describes the pump oil's function as "just lubrication — like motor oil in a car engine." Explain why this description is incomplete for a rotary vane vacuum pump.
Question 5 — Question 5
Which of the following pump types operates by transferring momentum to individual gas molecules using high-speed rotating blades?
Question 6 — Question 6
R1-A has been pumping at base pressure (0.05 mbar) for several hours with R1-V-ISO open and no gas flowing through the chamber. A thin-film coating specialist inspects the chamber and reports finding a faint oily film on the internal chamber walls. The system has no foreline trap. What is the most likely source of this contamination, and what is the mechanism?
Question 7 — Question 7
An operator observes the following on R1-A: the pump (R1-P-RP) sounds normal, the oil sight glass shows clear amber oil at the correct level, R1-FLT-EXH exhaust is clean (no visible mist), and the pump body is warm but not unusually hot. However, the system is taking 15 minutes to reach 0.05 mbar instead of the usual 8 minutes. A rate-of-rise test shows a decreasing rate from 0.03 to 0.005 mbar/min. What does this evidence tell you about the likely cause?
Question 8 — Question 8
R1-A has two filters — R1-FLT-VENT on the vent line and R1-FLT-EXH on the exhaust line. A colleague describes them both as "just filters" and questions why two are needed. Describe the distinct function of each filter and explain why each is necessary.
Question 9 — Question 9
On a more complex system than R1-A, a turbomolecular pump is backed by a rotary vane pump. The turbo pump is operating normally, but the backing pump's oil has become contaminated and its ultimate pressure has degraded from 0.01 mbar to 0.5 mbar. What effect would this have on the system?
Question 10 — Question 10
R1-A has been shut down for maintenance. When the technician restarts R1-P-RP, the pump makes a loud knocking sound for approximately 30 seconds, then settles into its normal operating sound. The pump reaches base pressure normally. Is this a concern?
Question 11 — Question 11
R1-A's maintenance log shows the following entries over three months:
Question 12 — Question 12
A colleague argues: "On R1-A, the pump-down is slow below 1 mbar. Since R1-P-RP only has a pumping speed of 5 m³/h, we should upgrade to a 20 m³/h pump to speed things up." The foreline (R1-L-FL) is 15 mm bore, 1.2 m long, with two 90-degree bends. Using M03 conductance concepts, interpret whether the proposed pump upgrade would achieve the expected improvement.
Question 13 — Question 13
On R1-A, after a normal pump-down to 0.05 mbar, the operator isolates the chamber (R1-V-ISO closed) and turns off R1-P-RP. The pressure begins to rise. After 10 minutes, R1-G-CH reads 0.15 mbar. After 60 minutes, it reads 0.85 mbar. After 6 hours, it reads 12 mbar. After 24 hours, it reads 47 mbar. The rates decrease significantly over time. Is this behaviour normal for an isolated rough-vacuum system?
Question 14 — Question 14
R1-A's pump oil is inspected and appears milky/cloudy instead of the normal clear amber. The system has been used in a workshop with very high humidity, and the chamber has been pumped down and vented repeatedly. What is the most likely explanation for the oil condition, and what is the consequence for pump performance?
Question 15 — Question 15
A vacuum system uses a scroll pump instead of a rotary vane pump for a semiconductor application. The scroll pump has a slightly higher ultimate pressure (0.01 mbar vs 0.001 mbar). A project manager asks: "Why are we paying more for a pump that performs worse?" Explain why the scroll pump is used in this application despite its higher ultimate pressure.
Question 16 — Question 16
A junior technician on a multi-pump system observes that the turbomolecular pump has been running for 10 minutes but the chamber pressure (reading on the high-vacuum gauge) has not dropped below 0.5 mbar. The roughing pump is running and its own inlet reads 0.8 mbar. What is the most likely problem, and which module's knowledge helps diagnose it?
Question 17 — Question 17
On R1-A, R1-P-RP is observed vibrating noticeably more than usual. The pump body is also hotter than normal. The oil sight glass shows oil at the correct level, but the oil appears slightly foamy. The system is reaching 0.08 mbar instead of the normal 0.05 mbar. Which of the following best describes a structured R-I-C-E observation (Recognise, Interpret, Communicate, Escalate) for this scenario?
Question 18 — Question 18
The following evidence is observed on R1-A:
Question 19 — Question 19
The following evidence is collected from R1-A for a capstone analysis integrating all modules. R1-A shows the following: base pressure of 0.15 mbar (normal: 0.05 mbar), pump-down from 950 to 1 mbar takes 90 seconds (normal), pump-down from 1 to 0.15 mbar takes 25 minutes and stalls (normal: 6.5 minutes to 0.05 mbar). Rate-of-rise: 0.06 mbar/min at 1 min, decreasing to 0.04 mbar/min at 10 min. Pump independent test: 0.009 mbar. Oil: slightly dark. R1-FLT-EXH: clean exhaust. Foreline: original specification (25 mm bore, 0.6 m, straight). No chamber modifications. All seals replaced 2 months ago. Identify the TWO most likely contributing factors and explain what discriminating evidence would separate them.
Question 20 — Question 20
Which of the following correctly describes the function of R1-FLT-EXH on R1-A?
Answer all questions to enable submission