Module 6

Assessment Content

Pumping Principles & Pumping Behaviour
Facilitator: adjust scaffolding level before distributing

S1 Entry Ticket: Pre-Synchronous Session Diagnostic (Capstone)

Scenario: "The Week After Maintenance"

R1-A was taken out of service last week for a comprehensive maintenance session. The following work was performed:

  1. R1-P-RP oil was changed (pump oil had been in service for 8 months)
  2. The foreline (R1-L-FL) was replaced because the original tubing had developed a dent. The replacement tubing was sourced from workshop stock: 16 mm internal diameter, 1.0 m long, with one 90-degree bend. The original foreline was 25 mm internal diameter, 0.6 m long, and straight.
  3. The chamber flange O-ring was replaced with a new Viton O-ring from sealed stock
  4. R1-FLT-EXH was replaced with a new filter element
  5. The chamber interior was wiped down with isopropanol cleaning solvent

After maintenance, the system was reassembled and returned to service on Monday morning. The first pump-down produces the following data:

Pump-down data:

Time from start R1-G-CH (mbar) Notes
0 min 950 Start from VENTED state
1.5 min 1.2 Viscous flow phase — slightly slower than the reference 90 seconds to 1.0 mbar
5 min 0.55 Below 1 mbar, pump-down is significantly slower than reference
10 min 0.35 Still dropping but slowly
20 min 0.22 Nearly stalled
30 min 0.20 Stalled — no further progress

Rate-of-rise test (isolated at 0.20 mbar, R1-V-ISO closed, R1-V-VENT closed, pump off):

Time after isolation R1-G-CH (mbar) Rate (mbar/min)
0 min 0.20
1 min 0.28 0.080
3 min 0.40 0.060
5 min 0.48 0.040
10 min 0.60 0.024

Additional evidence:

Entry Question 1: Phase Analysis (M03 Integration)

Compare the viscous flow phase to the molecular flow phase. Is only one phase affected, or are both? What does this comparison tell you about the location and nature of the problem?

Your answer:

Clue: Compare viscous and molecular flow phases to the reference data. Remember the conductance concepts from M03.
Guide: The viscous flow phase is slightly slow (1.5 min vs 1.5 min reference). The molecular flow phase is severely affected — stalling at 0.20 mbar instead of reaching 0.05 mbar in 8 minutes. This pattern should remind you of M03 (conductance restriction) but the rate-of-rise data adds another layer.

Entry Question 2: Rate-of-Rise Interpretation (M02 Integration)

Examine the rate-of-rise values in the table above. Is the rate constant or decreasing over time? What type of gas source does this pattern indicate — a leak or outgassing?

Given the maintenance activities performed, identify at least two specific sources that could produce this gas load pattern.

Your answer:

Clue: Remember the M02 diagnostic pattern. Is the rate constant (leak) or decreasing (outgassing)?
Guide: Rate decreases from 0.080 to 0.024 mbar/min — this is outgassing (decreasing/desorption pattern), not a leak. But the magnitude is high. What was recently introduced that could outgas significantly? Consider the maintenance history.

Entry Question 3: Pump Assessment (M06)

The pump reaches 0.008 mbar independently. The system stalls at 0.20 mbar.

What does this 25-fold gap between pump capability and system performance tell you? Is the pump a contributing factor to the problem?

Your answer:

Clue: The pump was tested independently at its own inlet. How does that compare to the system performance?
Guide: Pump alone: 0.008 mbar in 4 minutes (within spec). Through the system: stalls at 0.20 mbar. The gap proves the pump is capable — the limitation is between the pump and the chamber.

Entry Question 4: Foreline Analysis (M03 Integration)

The replacement foreline (16 mm bore, 1.0 m, one bend) replaced the original (25 mm bore, 0.6 m, straight). Without calculations, describe the effect of each change (diameter, length, bend) on conductance, and explain which flow regime is most affected. Could this foreline change alone explain the full performance degradation?

Your answer:

Clue: Compare the original foreline dimensions to the replacement. Remember the conductance concepts from M03.
Guide: Original: 25 mm bore, 0.6 m, straight. Replacement: 16 mm bore, 1.0 m, one bend. In molecular flow, conductance scales with d^3 and inversely with length. The diameter change alone reduces conductance to roughly (16/25)^3 = 26% of original.

Entry Question 5: Multi-Factor Diagnosis

This scenario likely has more than one contributing factor. Based on all the evidence, rank the following hypotheses and explain your reasoning:

For each hypothesis, state whether the evidence supports, contradicts, or is neutral. Identify at least one discriminating test for each.

Your answer:

Clue: This capstone scenario integrates M02 through M05. Which concepts from each module apply?
Guide: M02: rate-of-rise pattern identifies outgassing. M03: foreline geometry restricts molecular flow conductance. M04: O-ring material outgassing, solvent residue. M05: system layout determines where each effect manifests. Build your evidence chain linking specific data to specific module concepts.
Step by step
Full cross-module analysis: The rate-of-rise (M02) shows decreasing rate = outgassing, not leak. The pump-down phase split (M03) shows molecular flow affected more than viscous = conductance restriction. The O-ring and solvent (M04) are the primary gas sources. The foreline geometry (M03) limits how fast the pump can remove that gas. Both factors combine to stall the system at 0.20 mbar.

Entry Question 6: Capstone Escalation Note

Write a 3-sentence escalation note for the maintenance supervisor: (1) what was observed, (2) what the evidence indicates (including what has been ruled out and what contributing factors remain), and (3) what additional information would help clarify the situation (include at least 3 UNKNOWN items that require further evidence to resolve).

Your answer:

Clue: This is the capstone escalation — synthesize evidence from multiple modules into a professional diagnostic note.
Guide: Include: the stalled pressure (0.20 mbar vs 0.05 mbar reference), the rate-of-rise evidence (decreasing = outgassing), the two contributing factors (foreline conductance restriction + chamber contamination from solvent and new O-ring), and recommended actions (extended pump-down to clear contamination, then foreline replacement to restore molecular flow conductance).

S2 Worked Example: Entry Ticket Model Answer

Entry Question 1 — Model Answer

Both phases are affected, but to different degrees. The viscous flow phase is slightly slower than reference (1.2 mbar at 1.5 min vs the expected 1.0 mbar at 1.5 min) — this suggests a modest conductance reduction that affects even the high-pressure phase, consistent with the narrower replacement foreline. The molecular flow phase is severely degraded (stalled at 0.20 mbar vs the normal 0.05 mbar at 8 minutes).

The fact that both phases are affected points to a conductance issue (the foreline) that impacts all pressures. But the disproportionate impact on the molecular flow phase tells us there is an additional factor: in molecular flow, conductance depends on geometry alone (M03) and scales with the cube of diameter, making the narrower foreline far more restrictive.

Additionally, the molecular flow phase is where surface gas load dominates — and the cleaning solvent and new O-ring introduce extra gas load that the already-restricted foreline cannot remove efficiently. Multiple factors are likely compounding.

Entry Question 2 — Model Answer

The rate-of-rise values provided in the data table show a clearly decreasing pattern: 0.080, 0.060, 0.040, 0.024 mbar/min over the 10-minute observation. This is the outgassing (depleting source) pattern — not a constant-rate leak.

Given the maintenance activities, two specific gas load sources that could produce this pattern:

  1. Isopropanol solvent residue: The chamber was wiped with isopropanol and allowed only 10 minutes to "air dry." Isopropanol evaporates quickly at atmospheric pressure but can be absorbed into surface micro-porosity on stainless steel and into the new O-ring elastomer. Under vacuum, this trapped solvent outgasses as a significant gas load — particularly heavy initially (high vapour pressure), depleting over hours as the solvent inventory is exhausted.
  1. New Viton O-ring outgassing: Fresh elastomer O-rings contain absorbed atmospheric gases, moisture, and residual manufacturing compounds. A new O-ring outgasses more heavily than a conditioned one — the outgassing rate decreases as the ring is exposed to repeated pump-down cycles. This contributes a moderate, slowly depleting gas load.

Both sources are consistent with the decreasing rate-of-rise pattern and the maintenance history.

Entry Question 3 — Model Answer

The pump reaches 0.008 mbar at its own inlet — well within its rated specification. The system stalls at 0.20 mbar — 25 times higher.

This gap proves definitively that the pump is NOT the problem. The pump has ample capability that is not reaching the chamber.

The gap is attributable to two factors: (1) the replacement foreline's reduced conductance limits how much of the pump's speed is delivered to the chamber (M03 effective pumping speed), and (2) the elevated gas load from solvent residue and O-ring outgassing adds gas that the already-restricted effective pumping must overcome. The pump is healthy — the bottleneck is between the pump and the chamber.

Entry Question 4 — Model Answer

Diameter reduction (25 mm to 16 mm): In molecular flow, conductance scales approximately with the cube of the diameter. The 36% reduction in diameter (25 to 16 mm) reduces molecular flow conductance dramatically — a rough estimate suggests approximately 75% reduction from diameter alone. In viscous flow, the effect is present but less severe because viscous flow conductance also depends on pressure (higher pressure means higher conductance), partially offsetting the geometric restriction.

Length increase (0.6 m to 1.0 m): Conductance decreases inversely with length. The 67% increase in length reduces conductance by approximately 40% (independent of the diameter effect). Longer tubes mean more wall collisions for molecules, randomising their direction and increasing the fraction that bounce back toward the chamber.

Added bend (one 90-degree elbow): Each bend forces molecules to collide with the elbow wall, further randomising their direction. In molecular flow, where there is no bulk gas stream to carry molecules around corners, bends are particularly costly. The bend adds an effective length increase and a directional randomisation penalty.

Most affected regime: Molecular flow is most affected because conductance in molecular flow depends entirely on geometry. In viscous flow, the pressure-dependent component of conductance partially compensates for geometric restrictions.

Could the foreline alone explain the full degradation? Partially, but probably not entirely. The foreline reduction explains the slow molecular flow phase and the modest viscous phase delay. But the rate-of-rise test shows a high initial outgassing rate (0.080 mbar/min) that is above what geometry alone would produce — the solvent and O-ring gas loads are additional contributing factors.

If the foreline were the only problem, the rate-of-rise would be lower (the chamber's normal outgassing would be the only source). The elevated rate-of-rise points to additional gas load on top of the conductance restriction.

Entry Question 5 — Model Answer

Rank Hypothesis Verdict Key Evidence Discriminating Test
1 A: Foreline conductance reduction Supported (major contributor) Both viscous and molecular phases affected; foreline geometry significantly degraded (bore reduced 36%, length increased 67%, bend added). Pump reaches 0.008 mbar independently but system stalls at 0.20 mbar — the gap is conductance-related. If the original-specification foreline (25 mm bore, 0.6 m, straight) were restored and base pressure improved significantly, that would confirm conductance as a major factor.
2 B: Solvent residue outgassing Supported (contributing factor) Isopropanol cleaning with only 10 minutes air-dry time; decreasing rate-of-rise at 0.080 mbar/min initial rate is elevated above normal chamber outgassing; solvent trapped in surface micro-porosity would produce exactly this pattern. If multiple consecutive pump-down cycles each show a lower rate-of-rise and faster pump-down, the solvent source is depleting — confirming it was contributing. After 4–5 cycles, the solvent should be largely exhausted.
3 C: New O-ring outgassing Plausible (minor contributor) New elastomers outgas more than conditioned ones. The contribution is real but typically small compared to solvent residue. This gas load would decrease over the first several days of pump-down cycling. After several pump-down cycles, if the rate-of-rise decreases but does not return to normal, the O-ring's contribution may still be present. If the O-ring were substituted with a pre-conditioned one (pumped separately), that would isolate this factor.
4 D: Pump failure Eliminated Pump reaches 0.008 mbar independently — within specification. Oil is fresh and clean. No abnormal noise, temperature, or exhaust. All pump health indicators are normal. No further test needed — the independent pump test is conclusive.

Multi-factor conclusion: The most likely diagnosis is a combination of Hypotheses A and B as the dominant factors, with C as a minor contributor. The foreline conductance restriction limits how fast gas can be removed (a permanent geometry problem that will not improve with time). The solvent residue adds a large but temporary gas load.

Together, they compound: the restricted foreline cannot remove the elevated gas load efficiently, producing the stall at 0.20 mbar. As the solvent depletes over multiple cycles, performance will partially improve — but the conductance restriction will prevent full recovery until the foreline is replaced.

Entry Question 6 — Model Answer

"R1-A is stalling at 0.20 mbar following last week's maintenance (normal base: 0.05 mbar in 8 minutes); both viscous and molecular flow phases are slower than reference, R1-P-RP reaches 0.008 mbar independently (pump eliminated as a cause), and a rate-of-rise test shows a high but decreasing rate (0.080 to 0.024 mbar/min). The evidence indicates two contributing factors — the replacement foreline (16 mm bore, 1.0 m, one bend vs original 25 mm bore, 0.6 m, straight) has severely reduced molecular flow conductance, and solvent residue from the isopropanol cleaning (only 10 minutes air-dry time) is producing elevated outgassing — while pump failure and a true leak are eliminated by the independent pump test and the decreasing rate-of-rise pattern respectively. UNKNOWN: (1) whether original-specification foreline tubing is available in maintenance stock, (2) whether the solvent residue is confined to the chamber surfaces or has also been absorbed into the new O-ring elastomer (repeated cycling would reveal this), and (3) whether any other connections were disturbed during maintenance that might contribute a small additional leak masked by the outgassing — a follow-up rate-of-rise test after the solvent depletes would reveal any constant-rate component."

S3 Worked Example: Situation Report

Scenario Context (Facilitator-Provided During Synchronous Session)

R1-A has been returned to service after the foreline was restored to original specification. The solvent residue has been depleted through five pump-down cycles over two days. The system is now on its sixth pump-down.

R1-G-CH reads 0.07 mbar at 10 minutes into the pump-down and is still dropping slowly. The operator is uncertain whether the system is recovering as expected or whether a residual problem remains.

Model Situation Report

System: R1-A State: ROUGHING (R1-V-ISO open, R1-V-VENT closed, R1-P-RP running) Time: 14:20

Observation: R1-G-CH reads 0.07 mbar at 10 minutes into pump-down. R1-G-BX reads 949 mbar. Viscous flow phase completed in approximately 90 seconds (normal).

The pump-down below 1 mbar is slower than the historical 8-minute reference but significantly improved from the 30-minute stall at 0.20 mbar seen before the foreline replacement.

R1-P-RP sounds normal. R1-FLT-EXH exhaust is clean.

Oil sight glass shows clear amber oil at the correct level. No unusual odour.

Interpretation: The system is in the molecular flow regime at 0.07 mbar. The restored foreline (original 25 mm bore, 0.6 m, straight) has recovered most of the conductance that was lost. The remaining 0.02 mbar above normal base (0.07 vs 0.05 mbar) is consistent with the new O-ring still conditioning — fresh Viton outgasses slightly more than a conditioned ring.

Over the next several days of daily pump-down cycling, this should decrease toward normal. The five prior pump-down cycles appear to have depleted the bulk of the solvent residue (the system no longer stalls and the rate of approach to base pressure is close to normal).

Comparison to expected: Logbook reference: 0.05 mbar at 8 minutes. Today: 0.07 mbar at 10 minutes.

The system is approximately 85% recovered. The remaining gap is small and consistent with O-ring conditioning — not a new problem.

Unknowns — Evidence Needed:

  1. UNKNOWN: Will the system reach 0.05 mbar if given additional pump-down time? Continued monitoring over the next 10 minutes would clarify this.
  2. UNKNOWN: Is the rate of pressure decrease still declining (approaching base pressure), or has it levelled off (approaching a stall)? R1-G-CH readings at 2-minute intervals would distinguish between these two patterns.
  3. UNKNOWN: Has the new O-ring's outgassing rate decreased compared to the first pump-down cycle two days ago? A comparison of today's molecular flow pump-down rate to the Day 1 post-foreline-replacement data would indicate whether conditioning is progressing.
  4. UNKNOWN: If the system does not reach 0.05 mbar within 20 minutes, a rate-of-rise test would help check for any constant-rate component that might indicate a small leak masked by the earlier outgassing.

Escalation: "R1-A roughing in progress — sixth pump-down since foreline restoration. Currently at 0.07 mbar at 10 minutes, compared to the post-maintenance stall at 0.20 mbar. System is recovering as expected.

Slightly above the historical 0.05 mbar base — consistent with new O-ring conditioning. Monitoring to confirm base pressure recovery. If the system stalls above 0.06 mbar after 20 minutes, further investigation would be warranted."

S4 Worked Example: Evidence Brief

Scenario Context

Following the full investigation of R1-A's post-maintenance performance problem, all evidence has been collected and corrective actions have been taken. This evidence brief documents the complete investigation for the maintenance record.

Model Evidence Brief

System: R1-A State during investigation: ROUGHING (stalled at 0.20 mbar), then ISOLATED (rate-of-rise test) Investigation: Post-maintenance performance assessment — capstone multi-factor diagnosis

State Call: Initially ROUGHING — R1-V-ISO open, R1-V-VENT closed, R1-P-RP running. Then ISOLATED for rate-of-rise. Confirmed by valve positions and gauge readings.

Observed Evidence:

Plausibility Check: The pump reaches 0.008 mbar independently, confirming the pump is not the problem. The 25-fold gap between pump capability (0.008 mbar) and system performance (0.20 mbar) with no leak detected (decreasing rate-of-rise) points to two compounding factors: reduced conductance (the foreline) and elevated gas load (solvent residue). The viscous flow phase delay is consistent with a conductance restriction that affects all pressures, while the molecular flow phase is disproportionately affected (cube-of-diameter dependence in molecular flow + additional gas load).

Hypotheses (ranked):

Rank Hypothesis Supporting Evidence Contradicting Evidence Discriminator
1 Foreline conductance restriction (major contributor) Foreline bore reduced 36%, length increased 67%, bend added; both flow phases affected; pump independent test shows large gap None — all evidence supports If original-specification foreline were restored and base pressure improved, that would confirm this factor
2 Solvent residue outgassing (significant contributor) 10-minute air-dry is insufficient; high initial rate-of-rise (0.080 mbar/min) is elevated above normal; decreasing pattern matches volatile source Decreasing rate could also be normal O-ring outgassing (but the magnitude is too high for O-ring alone) If rate-of-rise decreases with each consecutive pump-down cycle, the solvent source is depleting
3 New O-ring outgassing (minor contributor) Fresh elastomers outgas more than conditioned ones; small but real contribution Magnitude too small to explain the full degradation alone Will resolve naturally over days of cycling
4 Pump failure None Pump reaches 0.008 mbar independently; all visual checks normal Eliminated
5 Seal leak None — no constant-rate component in the rate-of-rise Decreasing rate-of-rise contradicts leak hypothesis Eliminated (verify with post-solvent-depletion rate-of-rise)

Resolution:

  1. Foreline replaced with original specification (25 mm bore, 0.6 m, straight) — conductance restored
  2. Five pump-down cycles performed over two days — solvent residue depleted
  3. System now reaching 0.07 mbar at 10 minutes (sixth cycle) — recovering toward normal
  4. Expected full recovery to 0.05 mbar within several additional days as O-ring conditions

UNKNOWN — Evidence Needed:

  1. Whether the system will reach exactly 0.05 mbar or settle at a slightly elevated base — continued monitoring will confirm
  2. Whether the dented original foreline can be repaired or must be permanently replaced — investigate stock of 25 mm tubing
  3. Whether the 10-minute solvent air-dry procedure is sufficient or whether a longer drying time (or vacuum-assisted drying) would have prevented the elevated outgassing — this question remains open pending further evidence from the current recovery cycle

Escalation Note: "Post-maintenance investigation on R1-A complete.

Two primary causes identified: (1) replacement foreline with degraded geometry (16 mm bore, 1.0 m, one bend) caused a severe conductance bottleneck, and (2) insufficient solvent drying after isopropanol cleaning produced elevated outgassing. Pump verified healthy. No leak found.

Corrective actions taken: foreline restored to original specification, solvent depleted through repeated pump-down cycles. System is recovering — currently at 0.07 mbar (sixth cycle), expected to return to 0.05 mbar as the new O-ring conditions. It remains to be confirmed whether the maintenance SOP adequately specifies original foreline dimensions and minimum solvent drying time."

S5 Worked Example: Sector Lens Output

Scenario Context

Using the capstone multi-factor scenario, the student applies the thin-film coating sector lens. This is the final sector lens of the course, integrating all six modules.

Model Sector Lens Output

Base scenario: R1-A post-maintenance performance degradation — conductance bottleneck from replacement foreline compounded by solvent residue outgassing, stalling at 0.20 mbar (normal: 0.05 mbar).

Sector: Thin-Film Coating

Sector Lens Application:

In a thin-film coating production environment, this multi-factor failure would have cascading consequences:

Sector-Specific Escalation: "Post-maintenance investigation on R1-A identified two simultaneous failure modes with direct relevance to thin-film coating quality: (1) replacement foreline with non-specification geometry caused a conductance bottleneck preventing adequate base pressure, and (2) insufficient solvent drying introduced hydrocarbon outgassing into the chamber environment. In a coating application, either factor alone would be sufficient to produce defective films. Together, they would halt production entirely.

Corrective actions implemented: foreline restored to specification, solvent residue depleted. Questions that remain open: (a) whether the maintenance SOP adequately specifies exact component dimensions with sign-off requirements, (b) whether current solvent cleaning procedures specify sufficient drying time or a vacuum-assisted drying protocol, and (c) whether a pre-production qualification pump-down (with rate-of-rise verification) is specified after any maintenance that opens the chamber or replaces gas-path components."

S6 Reading List

Use these references to deepen your understanding of the concepts covered in Module 6. They are organised by topic and include section references for quick navigation.

Source Author/Publisher Topic Sections Priority Why Read This
Introduction to Vacuum Technology, Ch. 4-5 Milne Open Textbooks Vacuum pumps: operating principles, types, performance characteristics; pump selection for applications Ch. 4 (roughing pumps), Ch. 5 (high-vacuum pumps) Start here The clearest introductory treatment of pump operating principles. Non-mathematical, with diagrams of rotary vane, scroll, diaphragm, and turbo pump mechanisms. Builds directly on the flow regime concepts from Chapter 3 (M03). Use this as your primary reference for understanding how each pump type works.
Basic Vacuum Practice, Ch. 2-3, 6-8 Varian (3rd Edition) Roughing pump and high-vacuum pump mechanisms; pump performance curves, pumping speed, ultimate pressure, backing requirements, pump selection Ch. 2 (pp. 39-58), Ch. 3 (pp. 59-90), Ch. 6 (pp. 136-170), Ch. 7 (pp. 171-200), Ch. 8 (pp. 201-225) Core The most practical treatment of pump performance in a single source. Ch. 2-3 provide detailed cross-section diagrams of rotary vane, diffusion, turbo, and ion pump mechanisms. Ch. 6-8 cover pumping speed curves, effective speed calculations, and system integration. Strong on rotary vane pump oil management and backstreaming.
Vacuum Technology Book II, Part 2 Pfeiffer Vacuum Pump types, pumping speed, ultimate pressure, pump combinations, system design Sections 3.1-3.5 (pp. 39-72) Core Authoritative reference for pump specifications and performance data. Particularly strong on turbomolecular pump principles and backing pump requirements. Use alongside Varian for a complementary manufacturer perspective. Note: Pfeiffer uses sea-level atmospheric pressure (1013 mbar) — remember Selkirk atmospheric is ~950 mbar.
Introduction to Vacuum Science (KJLC/ORNL deck) J.R. Gaines, Kurt J. Lesker Company Pump mechanisms, performance curves, pump selection, system integration Slides 200-280 Recommended Excellent visual reference for pump cross-sections, performance curves, and system-level pump integration. The slide format makes pump mechanisms easy to visualise. Particularly useful for understanding the pumping sequence (rough-crossover-high vac) and backing requirements.
A User's Guide to Vacuum Technology, Ch. 5-7 John F. O'Hanlon Detailed treatment of roughing pumps, high-vacuum pumps, pump oil management, backstreaming, and contamination control Ch. 5-7 Supplementary More detailed than needed for Module 6's observational approach, but valuable if you want deeper understanding of pump internals, oil chemistry, backstreaming mechanisms, and contamination control strategies. Strong on the diagnostic relationship between pump health and system performance. Good reference for engineers in the cohort.

How to Use This List:

KJLC/ORNL Deck — Slide Guide for Module 6:

Lesson Slide Range What You'll Find
Lesson 2 (Roughing Pumps) 200-230 Cross-section diagrams of rotary vane, scroll, and diaphragm pump mechanisms; performance comparison tables
Lesson 3 (High-Vacuum Pumps) 231-255 Turbo pump rotor-stator diagrams; compression ratio concepts; backing pump requirements
Lesson 4 (Pump Performance) 256-270 Pumping speed curves; ultimate pressure concepts; effective speed at the chamber
Lesson 5 (Pump Safety) 271-280 Pump hazards overview; oil management; exhaust filtration; problematic behaviour recognition

End of Assessment Content — Module 6

Submit Your Assessment

Use the fields below to submit your completed assessment work. You may paste your Entry Ticket, Situation Report, Evidence Brief, or Sector Lens responses into the appropriate fields.