Module 2

Assessment Content

Influences on Real Vacuum Systems
Facilitator: adjust scaffolding level before distributing

§1 Entry Ticket: Pre-Synchronous Session Diagnostic

Scenario: "The Tuesday Afternoon Mystery"

R1-A was used for a successful demonstration this morning. The system reached its normal base pressure of 0.05 mbar with no issues. After the demo, the system was vented (controlled vent through R1-FLT-VENT, R1-V-VENT opened slowly then closed, chamber at ~950 mbar) and left in VENTED state (both valves closed, pump off) over lunch.

This afternoon, a different operator begins roughing. The pumpdown is normal from 950 to 1 mbar (~2 minutes).

Below 1 mbar, the pumpdown slows significantly. After 30 minutes, R1-G-CH reads 0.20 mbar and is still dropping slowly.

The operator isolates the system (R1-V-ISO closed, R1-V-VENT closed, pump off) and records the following rate-of-rise data:

Time after isolation R1-G-CH (mbar)
0 min 0.20
1 min 0.30
3 min 0.43
5 min 0.50
10 min 0.58

Additional context: The morning operator mentions that the demonstration involved placing several rubber test pieces inside the chamber — pieces that had been stored in a humid workshop area.

Entry Question 1: Gas Load Pattern

The rate of rise for each interval has been calculated below:

Interval Rise (mbar) Rate of rise (mbar/min)
0–1 min 0.10 0.100
1–3 min 0.13 0.065
3–5 min 0.07 0.035
5–10 min 0.08 0.016

Examine the rate-of-rise values. Is the rate constant or decreasing over time? What gas load source does this pattern indicate — a leak or outgassing?

Your answer:

Clue: Look at the rate-of-rise column. Is the rate staying the same over time, or is it changing? The pattern is the key diagnostic.
Guide: The rate drops from 0.100 to 0.016 mbar/min over 10 minutes. A constant rate means a leak (unlimited atmospheric source). A decreasing rate means outgassing (finite surface source being depleted). Which pattern do you see here?
Step by step
R-I-C-E applied: Recognise the decreasing rate (0.100 -> 0.065 -> 0.035 -> 0.016). Interpret: this is desorption/outgassing, not a leak. A leak would show a constant rate because atmosphere is an unlimited gas source. Communicate: the rate pattern, not just the magnitude, is the diagnostic key.

Entry Question 2: Contamination Hypothesis

Given the system history (morning demo, rubber test pieces from humid storage), what are the two most likely sources of the elevated gas load? Explain your reasoning for each.

Your answer:

Clue: Think about what was placed inside the chamber and where it came from. What two sources of extra gas might this introduce?
Guide: The rubber test pieces from humid storage carry absorbed water that outgasses under vacuum. Additionally, the chamber was vented over lunch, allowing water vapour from ambient air to re-adsorb onto all internal surfaces. Both contribute to the elevated gas load.

Entry Question 3: Why Not a Leak?

The rate of rise is relatively high (0.10 mbar/min initially). How do you know this is NOT a real leak, despite the elevated rate? What specific evidence from the data supports your conclusion?

Your answer:

Clue: Focus on the shape of the rate-of-rise curve, not the initial magnitude. What does a decreasing rate tell you about the gas source?
Guide: Despite the high initial rate (0.10 mbar/min), the rate is clearly decreasing — from 0.10 to 0.016 mbar/min. A real leak has a constant rate because atmosphere is an unlimited supply. The decreasing rate proves the source is finite (outgassing from surfaces), not replenished (leak from atmosphere).

Entry Question 4: Vent Filter Check

The system was vented through R1-FLT-VENT this morning. If the extended pumpdown were caused by particulate contamination from a degraded vent filter, how would the pumpdown behaviour be expected to differ from what is being observed? Name one distinguishing feature.

Your answer:

Clue: Think about how particle contamination would affect the system differently from molecular contamination like water vapour.
Guide: Particulate contamination creates additional surface area for water adsorption and can trap gas pockets (virtual leaks). These virtual leaks produce consistent rate-of-rise patterns across repeated pump-down cycles — unlike true outgassing, which improves with each cycle.

Entry Question 5: Escalation Note

Write a 3-sentence escalation note for the shift supervisor summarising: (1) what was observed, (2) what the evidence indicates, and (3) what additional information would help clarify the situation. Use specific readings and component IDs.

Your answer:

Clue: Structure: observation, evidence-based interpretation, then your ask. Use specific component IDs and readings.
Guide: Mention: R1-G-CH stalled at 0.20 mbar (normal base is 0.05 mbar). Rate-of-rise pattern is decreasing (0.10 -> 0.016 mbar/min). This indicates outgassing, not a leak. Likely source: humid-stored rubber test pieces. Ask: are the pieces still inside?

§2 Worked Example: Entry Ticket Model Answer

Entry Question 1 — Model Answer

Interval Rise (mbar) Rate (mbar/min)
0–1 min 0.10 0.10
1–3 min 0.13 0.065
3–5 min 0.07 0.035
5–10 min 0.08 0.016

The rate decreases from 0.10 to 0.016 mbar/min over 10 minutes. This is a concave curve — the classic outgassing (desorption) pattern. The gas load source is surface-related: water desorption and/or material outgassing, not a real leak.

Entry Question 2 — Model Answer

Source 1: Water desorption from rubber test pieces. Rubber stored in a humid workshop will have absorbed significant water vapour. When placed in vacuum, this water slowly releases (outgasses), creating a large gas load in the 1–0.01 mbar range. Rubber is particularly problematic because it has high permeability and absorbs water readily.

Source 2: General surface water re-adsorption. The chamber was vented and left at atmosphere over lunch (~2 hours). During this time, water vapour from the ambient air re-adsorbed onto all internal surfaces. This is normal but adds to the total gas load for the afternoon pumpdown.

Entry Question 3 — Model Answer

Despite the high initial rate (0.10 mbar/min), this is NOT a real leak because the rate is decreasing — from 0.10 to 0.016 mbar/min over 10 minutes. A real leak produces a constant rate because atmosphere is an unlimited gas source.

The decreasing rate is the definitive evidence: the gas source is being depleted (surface desorption), not replenished (leak from atmosphere). The shape of the curve, not the initial magnitude, is the diagnostic key.

Entry Question 4 — Model Answer

If particulate contamination from a degraded vent filter were the primary cause, the extended pumpdown would likely show different characteristics: the system might show relatively normal pumpdown on the first cycle, with the particle-related effects (increased surface area for water adsorption) becoming apparent primarily after venting and re-pumping. Additionally, particles would introduce a risk of virtual leaks (trapped gas pockets) that would produce identical rate-of-rise patterns across repeated cycles — unlike true outgassing, which improves with each cycle. The current pattern (no particles reported, first pumpdown of the afternoon showing extended time) is more consistent with water/outgassing contamination than particulate contamination.

Entry Question 5 — Model Answer

"R1-A pumpdown stalled at 0.20 mbar after 30 minutes of roughing — normal base is 0.05 mbar. Rate-of-rise test shows decreasing rate (0.10 → 0.016 mbar/min over 10 min), consistent with elevated outgassing rather than a leak.

Likely cause: water and material outgassing from humid-stored rubber test pieces placed in the chamber this morning. Additional information needed: confirmation of whether rubber test pieces are still in the chamber, and whether an extended pump-down cycle reduces the base pressure toward 0.05 mbar."

§3 Worked Example: Situation Report

Scenario Context (Facilitator-Provided During Synchronous Session)

R1-A has been running for two hours. It initially reached 0.08 mbar (slightly above normal). Over the last 30 minutes in ROUGHING state, R1-G-CH has been gradually rising from 0.08 to 0.15 mbar while the pump continues running.

Model Situation Report

System: R1-A State: ROUGHING (R1-V-ISO open, R1-V-VENT closed, R1-P-RP running) Time: 14:30

Observation: R1-G-CH has risen from 0.08 mbar to 0.15 mbar over 30 minutes during active roughing. Pump is running and sounds normal. R1-G-BX reads ~950 mbar.

Interpretation: Pressure rising during active pumping means gas load is increasing — the pump can no longer maintain the initial base pressure. This is not normal during steady-state roughing. Possible causes: (1) developing leak at a seal (gas ingress increasing over time as the seal degrades or shifts); (2) thermal outgassing from a component that is warming up during extended operation; (3) pump oil backstreaming increasing at sustained low pressure.

Unknowns → Evidence Needed:

  1. UNKNOWN: Is the pressure rise constant or accelerating? → Record R1-G-CH every 5 minutes for the next 20 minutes to characterise the trend.
  2. UNKNOWN: Is this a leak or an outgassing/contamination source? → Isolate the system and perform a rate-of-rise test to check for constant vs decreasing rate.
  3. UNKNOWN: Has anything changed in the pump exhaust? → Check R1-FLT-EXH for unusual odour or oil discharge that might indicate pump issues.

Escalation: "R1-G-CH rising during active roughing — 0.08 → 0.15 mbar over 30 min. Pump running normally.

A rate-of-rise test after isolation would help determine whether the gas source is a developing leak or thermal outgassing. Will report findings after test."

§4 Worked Example: Evidence Brief

Scenario Context

Following the situation report above, the operator isolates R1-A. Rate-of-rise test results:

Time (min) R1-G-CH (mbar) Rate (mbar/min)
0 0.15
1 0.27 0.12
3 0.51 0.12
5 0.75 0.12
10 1.35 0.12

Model Evidence Brief

System: R1-A State during test: ISOLATED (R1-V-ISO closed, R1-V-VENT closed, R1-P-RP off) Test: Rate-of-rise, 10-minute duration

State Call: ISOLATED — both valves closed, pump off. Confirmed by valve positions.

Observed Evidence:

Plausibility Check: Constant rate is NOT consistent with outgassing (which should decrease). Constant rate IS consistent with a real leak — atmosphere providing a steady gas source at 0.12 mbar/min.

Hypotheses (ranked):

  1. Real leak (most likely) — constant rate from unlimited atmospheric source. Possible locations: R1-V-ISO seat (valve was recently operated), chamber O-ring (sealing surface), any flanged connection.
  2. Unusual constant outgassing source (less likely) — a large contamination source could produce a near-constant rate early in the test, but the absolute constancy over 10 minutes argues against this.

Discriminator Evidence:

UNKNOWN → Evidence Needed:

  1. Leak location — need physical inspection of seal surfaces, potentially helium leak testing
  2. Whether the leak developed gradually or suddenly — need operational log review
  3. Whether pump performance is contributing — need to verify pump ultimate pressure independently

Escalation Note: "Rate-of-rise test confirms constant gas ingress at 0.12 mbar/min — consistent with a real leak, not outgassing. The most likely leak locations are R1-V-ISO seal and the chamber O-ring. Additional information needed: physical inspection of seal surfaces and potentially helium leak testing to confirm the leak location."

§5 Worked Example: Sector Lens Output

Scenario Context

Using the leak scenario from §4, the student applies the General Industrial sector lens.

Model Sector Lens Output

Base scenario: R1-A real leak at 0.12 mbar/min, identified through rate-of-rise test.

Sector: General Industrial

Sector Lens Application:

In a general industrial vacuum setting (e.g., vacuum packaging, degassing, drying), this leak rate would mean:

Escalation (sector-specific): "Leak rate of 0.12 mbar/min identified on R1-A. In a general industrial vacuum application, this would prevent the system from meeting typical process specifications below 1 mbar.

Additional information needed: leak location confirmation through physical inspection or helium leak testing. Product quality is at risk if the system is used in this condition."

§6 Reading List

Use these references to deepen your understanding of the concepts covered in Module 2. They're organized by topic and include page numbers for quick navigation.

Source Author/Publisher Topic Sections Priority Why Read This
Introduction to Vacuum Technology, Ch. 2 Milne Open Textbooks Gas interactions with surfaces; outgassing; permeation Ch. 2 Start here Clear, accessible treatment of why gas sticks to surfaces and how it comes off under vacuum. Builds directly on Chapter 1 from Module 1.
Basic Vacuum Practice, Ch. 2–3 Varian (3rd Edition) Gas sources; leak detection; contamination Ch. 2 (pp. 36–60), Ch. 3 (pp. 61–85) Core The clearest explanation of gas load contributors, rate-of-rise testing, and leak classification available in a single source. Practical, plain-language approach.
Vacuum Technology Book II, Part 2 Pfeiffer Vacuum Leak detection methods; gas analysis; RGA fundamentals Sections 4.1–4.3 (pp. 45–58) Core Authoritative technical reference on leak detection and gas analysis. Excellent diagrams of leak types and RGA spectra.
Introduction to Vacuum Science (KJLC/ORNL deck) J.R. Gaines, Kurt J. Lesker Company Outgassing data; material properties; bake-out procedures Slides 250–320 Recommended Outstanding visual reference for outgassing rates of common materials, with real measurement data and practical bake-out guidance.
A User's Guide to Vacuum Technology, Ch. 6 John F. O'Hanlon Rate-of-rise testing; system diagnostics; gas load measurement Ch. 6 Supplementary Detailed treatment of diagnostic procedures. More advanced than needed for Module 2, but valuable if you want to go deeper into quantitative leak detection.

How to Use This List:

KJLC/ORNL Deck — Slide Guide for Module 2:

Lesson Slide Range What You'll Find
Lesson 2 (Gas Load Sources) 250–275 Outgassing rate data for common materials (metals, polymers, elastomers)
Lesson 3 (Rate of Rise) 276–295 Leak detection principles, rate-of-rise interpretation examples
Lesson 4 (Contamination) 296–310 Clean handling procedures, contamination sources, bake-out data
Lesson 5 (RGA & Diagnostics) 311–320 Mass spectrum examples, common gas species identification

End of Assessment Content — Module 2

Submit Your Assessment

Use the fields below to submit your completed assessment work. You may paste your Entry Ticket, Situation Report, Evidence Brief, or Sector Lens responses into the appropriate fields.