R1-A Scenario Animator
Module 1
System States
Idle / Resting
Both valves closed, pump OFF, ~950 mbar
Roughing
ISO open, VENT closed, pump ON, pressure dropping
Isolated
Both valves closed, watching for leaks
Vented (at atmosphere)
VENT open, ISO closed, chamber at atmosphere
Controlled Vent (transition)
VENT open, pressure rising toward ~950 mbar
Diagnostic Scenarios
Leak Detected
Isolated, pressure rising fast (3 mbar/min)
Base Pressure Reached
Pump at ultimate vacuum (~10⁻² mbar)
Error Scenarios
⚠ Vent Open During Pumping
Pump fights atmosphere — no vacuum possible
Guided Walkthrough
ATM
~950 mbar
ATM
R1-L-VENT
R1-L-FL
R1-L-EXH
R1-FLT-VENT
VENT
(angle valve)
R1-V-VENT
CHAMBER
R1-CH
ISO
(ball valve)
R1-V-ISO
PUMP
R1-P-RP
R1-FLT-EXH
R1-G-CH
950 mbar
R1-G-BX
~950 mbar
Chamber Pressure
950 mbar
10⁻³
1
100
950
LEGEND
Open / ON
Closed / OFF
Valve
Filter
Gauge
Roughing flow
Vent flow
IDLE
Idle / Resting State
System at local atmospheric pressure. Both valves closed. Pump is OFF. Chamber is at ~950 mbar. This is the safe resting state of the rig.
Component Status
Component
State