R1-A: Single-Stage Roughing Rig — P&ID (ISO 14617-2)
R1-L-VENT
R1-L-FL
R1-L-EXH
gas flow
vent flow
ATM
~950 mbar
R1-FLT-VENT
CLOSED
R1-V-VENT
R1-CH
~950 mbar
PI
Pirani
R1-G-CH
CLOSED
R1-V-ISO
R1-P-RP
R1-P-RP
OFF
R1-FLT-EXH
ATM
PI
baro
R1-G-BX
~950 mbar
Wall-mounted
ISO 14617-2 Symbol Key
Ball valve (filled circle = ball element)
Angle valve (chevron = 90-degree flow path)
Gate valve (plain bowtie) [R2-A/R3-A]
Pump (circle + rotation)
PI
Instrument (PI = Pressure Indicator)
Filter (diamond + screen line, ISO 14617-2)
VENTED — System at Atmospheric Pressure
All valves closed. Pump OFF. Chamber at ~950 mbar (Selkirk elevation). Safe resting state.
R1-A Overview
ISO 14617-2 P&ID
Components 6 core + 2 inline
Pressure Range ATM (~950 mbar) to ~10⁻² mbar
Primary Flanges KF25 (NW25)
Gauge Flange KF16 (via reducer)
System States 3 + Controlled Vent
Click any component for detailed specifications. Use the state buttons above to see valve positions and flow paths for each operating mode.
ISO 14617-2 defines graphical symbols for process diagrams. This P&ID uses standard valve, pump, instrument, and filter symbols as specified in the standard.
ROUGHING:
R1-CH → R1-V-ISO → R1-L-FL → R1-P-RP → R1-FLT-EXH → ATM
VENT:
ATM → R1-FLT-VENT → R1-L-VENT → R1-V-VENT → R1-CH